Measuring Power
High Performance Power Meter
49
Measuring Corrected Power
To set a correction factor:
1.
From the
Settings
function tab, select the
Channels
tab.
2.
Select the channel for which you want to set the CF (in the case of a
two- or four-channel power meter).
3.
Select the wavelength to which the CF will be applied by using the up
and down arrow buttons next to the list or by clicking it once in the
Correction Factors
list.
4.
In the
Factor (unit)
box, enter a CF between
10.000 dB and
10.000 dB, or between 0.100 W/W and 10.000 W/W, then click
.
Note:
Although the wavelength list applies to all channels, the CF applies to the
wavelength and channel at which it was set.
To remove the offset from a wavelength, select the wavelength,
then click
.