SCPI Command Reference
168
IQS-1700
Product-Specific Commands—Description
:SENSe[1..n]:POWer[:DC]:REFerence:
STATe?
Description
This query returns a value indicating whether the
power meter is displaying absolute (dBm or W)
or relative (dB or W/W) power values for the
specified channel.
Syntax
:SENSe[1..n]:POWer[:DC]:REFerence:STATe?
Parameter(s)
None
Response Syntax
<ReferenceState>
Response(s)
ReferenceState:
The response data syntax for <ReferenceState>
is defined as a <CHARACTER RESPONSE DATA>
element.
Current reference mode, where:
0 - is the absolute mode(dBm or W).
1 - is the relative mode (dB or W/W).
Example(s)
SENS:POW:REF:STAT?
See Also
SENSe:POWer:[DC]:REFerence
SENSe:POWer:[DC]:REFerence?
SENSe:POWer:[DC]:REFerence:ALL
SENSe:POWer:[DC]:REFerence:DISPlay
SENSe:POWer:[DC]:REFerence:STATe
Summary of Contents for IQS-1700
Page 1: ...User Guide High Performance Power Meter and Optical Head for IQS 600 IQS 1700...
Page 5: ...Contents High Performance Power Meter v Index 189...
Page 8: ......
Page 14: ......
Page 16: ......
Page 44: ......
Page 64: ......
Page 96: ......
Page 114: ......
Page 196: ......