
SCPI Command Reference
146
IQS-1700
Product-Specific Commands—Description
:SENSe[1..n]:CORRection:OFFSet
[:MAGNitude]
Description
This command sets an offset value. The units are
w/w by dÈfault.
If no channel was specified, the default channel
used is 1.
Syntax
:SENSe[1..n]:CORRection:OFFSet[:MAGNitude]
<wsp><CorrectionOffset[<wsp>W/W|DB]>|
MAXimum|MINimum|DEFault
Parameter(s)
CorrectionOffset:
The program data syntax for <CorrectionOffset>
is defined as a <numeric_value> element
followed by an optional <SUFFIX PROGRAM
DATA> element. The allowed <SUFFIX
PROGRAM DATA> elements are: W/W|DB. The
<CorrectionOffset> special forms MINimum,
MAXimum and DEFault are accepted on input.
MINimum allows to set the instrument to the
smallest supported value.
MAXimum allows to set the instrument to the
greatest supported value.
Summary of Contents for IQS-1700
Page 1: ...User Guide High Performance Power Meter and Optical Head for IQS 600 IQS 1700...
Page 5: ...Contents High Performance Power Meter v Index 189...
Page 8: ......
Page 14: ......
Page 16: ......
Page 44: ......
Page 64: ......
Page 96: ......
Page 114: ......
Page 196: ......