SCPI Command Reference
High Performance Power Meter
153
Product-Specific Commands—Description
:SENSe[1..n]:FREQuency:NCONtinuous
Description
This command sets the single acquisition rate in
Hz.
Syntax
:SENSe[1..n]:FREQuency:NCONtinuous<wsp><
SingleRate[<wsp>HZ]>
Parameter(s)
SingleRate:
The program data syntax for <SingleRate> is
defined as a <DECIMAL NUMERIC PROGRAM
DATA> element followed by an optional
<SUFFIX PROGRAM DATA> element. The
allowed <SUFFIX PROGRAM DATA> element is
HZ.
The <numeric_value> parameter is the Single
acquisition rate.
Example(s)
SENS:FREQ:NCON 512
Notes
This command is not executed if a data
acquisition is in progress. In that case, the
"acquisition in progress" message will be
returned.
See Also
SENSe:FREQuency:CONTinuous
SENSe:FREQuency:CONTinuous?
SENSe:FREQuency:CONTinuous:CATalog?
SENSe:FREQuency:NCONtinuous?
SENSe:FREQuency:NCONtinuous:CATalog?