Tunable Laser
Source
FLS-2800
FLS-2800 (v1.1)
21
General Parameters (changing frequency or wavelength)
Using the general tab in the Laser Parameter Control panel, the power and frequency or wavelength
(depending on the
Environment Settings
) can be set for the selected laser sources.
Figure 15 - Laser Parameters Single Laser
Figure 16 - Laser Parameters Multiple Laser Selection
When a single laser source is selected you can enter the desired values and click ‘
Set
’. The laser
source
Status Symbol
will reflect the progress of the ‘
Set
’ request.
If both laser sources are selected, a
Start
value for the power, frequency or wavelength can be
entered along with a
Step
size value. When you click ‘Set’ to apply parameter changes, the first
selected laser source will be set at the
Start
value. The remaining laser source will be set to value
Start + (Step)
.
Note
: The software will provide a warning when you attempt to ‘Set’ values that are not valid or out
of the operating range of the laser sources.