CableSHARK P3
VF/DSL Cable Qualifier
Figure 5.2.2A - Examples of Circuit where Single Ended Measurements can be used
5.2.3 End-to-End Testing
End-to-End testing allows two CableSHARK’s to connect to the same line to perform a variety of tests (as
shown in Figure 5.2.3A). Tests include End-to-End Frequency Response, DMT tests, and Load Coil
Detection. A CableSHARK at the Far End is the Tx unit, which is running in Remote Responder Mode. It
responds to all End-to-End test requests from the Rx unit as a controller and sends signals for the requested
test down the line to the Rx unit, which in turn converts the signal and analyzes the results. The Rx unit is
the one that analyzes and displays the information.
+ -
CableSHARK P3
xDSL Cable Qualifie r
F1
F2
F3
F4
F5
F6
F7
F8
1
2
A
B
C
3
D
E
F
4
G
H
I
5
J
K
L
6
M
N
O
7
P
R
S
8
T
U
V
9
W
X
Y
0
Q
Z
*
#
R U N
S T O P
E
F
CH G
A Wo rld of G oo d fo r Yo ur Ne two rk s
SYSTEM
SETU P
CABLE
TESTS
DMM
TESTS
RECALL
RESULTS
13 :16 :20
RX unit
(initiating test)
+ -
CableSHARK P3
xDSL Cable Qualifie r
F1
F2
F3
F4
F5
F6
F7
F8
1
2
A
B
C
3
D
E
F
4
G
H
I
5
J
K
L
6
M
N
O
7
P
R
S
8
T
U
V
9
W
X
Y
0
Q
Z
*
#
R U N
S T O P
E
F
CH G
MAIN
MENU
13 :16 :20
F
TX unit
Re spo nde r is waitin g f or a r emote -t est task. ..
Figure 5.2.3A- Example of a End to End Measurement
5.2.4 High Voltage Detection
The CableSHARK will provide the user with visual feedback on the CableSHARK’s LCD if a voltage over
60VAC or 60 VDC is detected. This message will appear only once a test has been executed and is meant
to protect the unit from damage and highlight to the user that a high voltage exists on the circuit under test.
Summary of Contents for CableSHARK P3
Page 1: ......
Page 10: ...VF DSL Cable Qualifier viii...
Page 12: ...VF DSL Cable Qualifier x...
Page 26: ...VF DSL Cable Qualifier 14...
Page 36: ...VF DSL Cable Qualifier 24...
Page 64: ...VF DSL Cable Qualifier 52...
Page 80: ...VF DSL Cable Qualifier 68...
Page 184: ...VF DSL Cable Qualifier 172...
Page 192: ...VF DSL Cable Qualifier 180...
Page 224: ...VF DSL Cable Qualifier 212...
Page 228: ...VF DSL Cable Qualifier 216...
Page 262: ...VF DSL Cable Qualifier 250 CSA Loop 6 CSA Loop 7 CSA Loop 8 Extended CSA Loop 9...
Page 263: ...CableSHARK P3 User Guide 251 Extended CSA Loop 10 Mid CSA Loop 0 Mid CSA Loop 1 Mid CSA Loop 2...
Page 264: ...VF DSL Cable Qualifier 252 Mid CSA Loop 3 Mid CSA Loop 4 Mid CSA Loop 5 Mid CSA Loop 6...
Page 265: ...CableSHARK P3 User Guide 253 ANSI Loop 2 ANSI Loop 3 ANSI Loop 4 ANSI Loop 5...
Page 266: ...VF DSL Cable Qualifier 254 ANSI Loop 6 ANSI Loop 7 ANSI Loop 8 ANSI Loop 9...
Page 267: ...CableSHARK P3 User Guide 255 ANSI Loop 11 ANSI Loop 12 ANSI Loop 13 ANSI Loop 15...
Page 268: ...VF DSL Cable Qualifier 256...
Page 286: ...VF DSL Cable Qualifier 274...
Page 292: ...VF DSL Cable Qualifier 280...
Page 294: ...VF DSL Cable Qualifier 282...
Page 316: ...VF DSL Cable Qualifier 304...
Page 318: ...VF DSL Cable Qualifier 306...
Page 319: ...CableSHARK P3 User Guide 307 Notes...
Page 320: ...VF DSL Cable Qualifier 308...