64
WT.090.370.000.DE.IM.0715
Operation
Barrier M
5.
**Analog inputs**
There are four analog inputs:
•
Water temperature Pt100
•
Transmission
•
Flow
•
UV signal
The current values of the analog inputs can be viewed. The values
for transmission, flow and UV signal are always between 4 and 20
mA, the value for the Pt100 water temperature is between 0 and
20 mA.
Press the acknowledge key to access the analog inputs submenus.
Current value of the analog input AD1 for the Pt100 water tempe-
rature
Legend of the analog input AD1
Current value of the analog input AD2 for the transmission
Legend of the analog input AD2
Current value of the analog input AD3 for the flow
Legend of the analog input AD3
Current value of the analog input AD4 for the UV signal
Legend of the analog input AD4
Press the UP key to return to the „Diagnosis” menu.
** Analog In. **
AD1 xxxx/ xx.xx mA
AD1=Pt100
AD2 xxxx/ xx.xx mA
AD2=Transmission
AD3xxxx/ xx.xx mA
AD3=Flow
AD4 xxxx/ xx.xx mA
AD4=UV-Signal
Summary of Contents for WALLACE & TIERNAN Barrier M 1200
Page 1: ...WALLACE TIERNAN UV DISINFECTION UNIT BARRIER M OPERATING MANUAL...
Page 2: ...2 WT 090 370 000 DE IM 0715 Barrier M Please note Original manual...
Page 96: ...96 WT 090 370 000 DE IM 0715 Spare parts maintenance schedule Barrier M 6...
Page 100: ...100 WT 090 370 000 DE IM 0715 Table of UV alarm levels Barrier M 7...
Page 102: ...102 WT 090 370 000 DE IM 0715 Certificates declarations and statements Barrier M 8...
Page 103: ...Barrier M Certificates declarations and statements 8 103 8 2 QPS certificate...
Page 104: ...104 WT 090 370 000 DE IM 0715 Certificates declarations and statements Barrier M 8...
Page 105: ...Barrier M Certificates declarations and statements 8 105 8 3 NSF certficate...
Page 106: ...106 WT 090 370 000 DE IM 0715 Certificates declarations and statements Barrier M 8...
Page 109: ...Barrier M 109...