Preprogrammed Settings
The monitoring system is preset to common Federal Drug Administration (FDA) and Center
for Disease Control (CDC) industry standards for pharmaceutical drug and vaccine storage
conditions. Settings can be customized via the Settings and Alerts menus.
Temperature Display & Alarm – The display is preset to read the glycol sensor but
can be changed to read the air sensor. Temperatures are detected and displayed in
Celsius units to one tenth of a degree. For example, 4.0 °C. Upper and lower alarm
thresholds are driven by the sensor selected for the display.
o
Set Point Temperature - The unit is preset to 4.0 °C/39.2 °F
o
Upper Threshold Alarm - preset to 8.0°C/46.4 °F
o
Lower Threshold Alarm - preset to 2.0°C/35.6 °F
SilverPoint Local Data Collection – The unit is programmed to record a temperature
reading every 15 minutes, with a capacity of 4000 data points.
Preset alarms/alerts
The following alarms/alerts are factory preset:
Set Point Temperature
When you first turn on the unit, the fan will run and the refrigerator will
work to reach the default set point temperature of 4 °C. Once set point is
reached, the fan speed will reduce to conserve energy. The unit will
reach the air temperature setting rapidly. Reaching Glycol set point
temperature will take longer.
Initially, the temperature range alarms are disabled until the unit reaches
its preset temperature range of 4 °C.