Aquafilter
E9_Next_Ing_rev03 – 10/10/2018
79
E9 NEXT
E44
Faulty condensation
battery temperature
sensor.
Fault in the sensitive
element or in the
probe connection.
Switch the steriliser off and then back on
again. If the problem persists contact the
assistance service.
E45
Chamber temperature
sensor faulty.
Fault in the sensitive
element or in the
probe connection.
Switch the steriliser off and then back on
again. If the problem persists contact the
assistance service.
E46
Pressure sensor
faulty.
Fault in the sensitive
element or in the
probe connection.
Switch the steriliser off and then back on
again. If the problem persists contact the
assistance service.
E47
Door closed sensor
faulty.
Fault in the door
closed position switch.
Open and close the door a few times. If the
problem persists contact the assistance
service.
E48
Door lock sensor
faulty.
Fault in the door
locked position switch.
Perform a sterilisation cycle. If the problem
persists contact the assistance service.
E51
Steam generator
inactive.
Fault in the steam
generator or electronic
board or steam
generator safety
thermostat tripped.
Contact the assistance service.
E54
Temperature of
steam generator too
high.
Fault in the electronic
board or temperature
sensor.
Switch off the steriliser and allow it to cool
for a few hours, then perform a sterilisation
system. If the problem persists contact the
assistance service.
E55
Temperature of upper
band too high.
Fault in the electronic
board or temperature
sensor.
Switch off the steriliser and allow it to cool
for a few hours, then perform a sterilisation
system. If the problem persists contact the
assistance service.
E56
Temperature of lower
band too high.
Fault in the electronic
board or temperature
sensor.
Switch off the steriliser and allow it to cool
for a few hours, then perform a sterilisation
system. If the problem persists contact the
assistance service.
E58
Temperature of lower
band heater too high
in the pressure
maintenance cycle.
Fault in the electronic
board or temperature
sensor.
Switch off the steriliser and allow it to cool
for a few hours, then perform a sterilisation
system. If the problem persists contact the
assistance service.
E59
Condensation battery
temperature too high
Sensor, connection, or
fan fault.
Contact the assistance service.
E60
Problems about write
on Sd card
SD card inserted after
power on or removed
during a cycle
Switch off the autoclave and check if the
sd card is in the right place. If the problem
persists contact the assistance service
E62
Water
injections
finished.
Boiler overloaded or
clogged with scale or
water injection pump
inefficient.
Perform a sterilisation cycle with a small
load in the boiler (just one tray). If the
problem persists contact the assistance
service.
E81
Water not being
supplied by the
Aquafilter® deioniser.
Error in the hydraulic
or electrical
connections with
Aquafilter or electronic
fault.
Make sure that the Aquafilter connections
are correct and that there are no crushed
or bent pipes. Make sure the Aquafilter
®
inlet tap is open. If the problem persists
contact the assistance service.
E99
Comunication
problem between
display board and
power board
Switch off the steriliser and after a few
seconds turn it on again. If the problem
persist contact the assistance service