Aquafilter
E9_Next_Ing_rev03 – 10/10/2018
63
E9 NEXT
9.3.1 Rusting
The formation of rust on the surfaces of the unit or instruments is caused by the introduction of rusty
instruments, even if made from stainless steel, or of instruments in normal steel that cause galvanisation to
take place.
The introduction of a single instrument with a rust stain is often sufficient to form and develop rust on the
instruments and in the unit itself.
Another factor that causes the formation of rust in the steriliser is the use of water containing chlorine, such
as drinking water, sea water or water containing disinfectants or detergents.
Only use good quality
distilled water.
CAUTION: DISCONNECT THE POWER SUPPLY BEFORE STARTING WORK. Non-
observance may cause serious injury to people or may seriously damage the unit.
If rust forms in the unit, clean the walls of the sterilization chamber and the tray holder using special products
for stainless steel, as described previously in the paragraph “Cleaning the sterilization chamber, accessories,
door and gasket”.
CAUTION:
do not use metal sponges or brushes. Use a damp soft cloth to remove dirt stains.