Aquafilter
E9_Next_Ing_rev03 – 10/10/2018
77
E9 NEXT
The table below lists all the alarm messages with the probable causes of faults; if your steriliser shows any of
the following error codes, before contacting the technical support service perform the checks indicated in the
table.
CODE
DESCRIPTION
PROBABLE CAUSE
SOLUTION
E01
Anomalous change in
input voltage.
Fault in the mains
power supply or
inadequate input
socket.
Make sure the unit is connected to a
suitable mains power supply.
E02
Black-out.
1.
Temporary black-
out.
2.
Two-pole thermal
safety switch
tripped.
3.
Safety thermostat
tripped.
1.
Wait for mains voltage to return.
2.
Switch the unit back on. If the problem
persists contact the assistance
service.
3.
Allow the steriliser to cool for a few
hours and then reset the safety
thermostat at the front of the machine.
If the problem persists contact the
assistance service.
E19
Excessive pressure
during sterilisation in
emergency cycles.
Fault during the
sterilization phase.
Allow the steriliser to cool and then
perform a sterilisation cycle with a small
load in the boiler (just one tray). If the
problem persists contact the assistance
service.
E21
Excessive pressure
during sterilisation.
Fault during the
sterilization phase.
Allow the steriliser to cool and then
perform a sterilisation cycle with a small
load in the boiler (just one tray). If the
problem persists contact the assistance
service.
E22
Insufficient pressure
during sterilisation.
Sterilisation phase
fault, or overload, or
steam loss.
Allow the steriliser to cool and then
perform a sterilisation cycle with a small
load in the boiler (just one tray). Make sure
no water leaks or drips on the front. If the
problem persists contact the assistance
service.
E23
Excessive
temperature during
sterilisation.
Fault during the
sterilization phase.
Allow the steriliser to cool and then
perform a sterilisation cycle with a small
load in the boiler (just one tray). If the
problem persists contact the assistance
service.
E24
Insufficient
temperature during
sterilisation.
Fault during the
sterilisation phase,
probably due to a leak
during a vacuum
phase.
Allow the steriliser to cool and then
perform a vacuum test. If this is successful,
perform a sterilisation cycle with a small
load in the boiler (just one tray). If the
problem persists contact the assistance
service.