Aquafilter
E9_Next_Ing_rev03 – 10/10/2018
32
E9 NEXT
When the cycle is completed normally, the sterile load is dried, and the door is released. The following
screen confirms that the process has been completed.
CAUTION:
when the sterilizer is turned off, be sure that the door is either open (
a
) or completely
closed (
b
). It is important to avoid the situation shown in
c
, which is closing the door with handle
not completely hooked.
Fig. 7.2.1-1
CAUTION
:
DANGER OF BURNS.
When the unit finishes the sterilization cycle and the door is
opened to remove the sterilized instruments, the inner parts of the boiler and door are still very
hot. These must not be touched directly in order to avoid getting burnt (chap. 3.4 “Residual
risks”). Use the relative extractor tool.
CAUTION
:
DANGER OF BURNS.
Do not lean over or stand in front of the door when opening it
as there is a risk of scalding from escaping steam (chap. 3.4 “Residual risks”). Use the relative
extractor tool.
If the sterilization cycle has not been successful, an error message will be displayed indicating the cause of
the problem (
Appendix 9 “Troubleshooting”
).