MANUAL SGM800
74
In
PcL 1
you can set the
polaritiy of the input range
. Un=Unipolar mode the input range
for load cells is –0.2 mV/V to + value set at
Range
. Bi=Bipolar mode the input range for
load cells is –value set at
Range
to + value set at
Range
.Press key 2 <2 sec. To change
the polaritity og the input range.
The following screen is visible:
Use key 1 change the polarity. Confirm by pressing key 3 for >2 sec.
Up
Down Confirm
The following screen is visible:
In
PcL 2
you can set the
Amplifier sensitivity
. Choose between 1.0mV/V, 1.5mV/ V,
2.0mV/V, 2.5mV/V and 3.0mV/V. Press key 2 <2 sec. to change the amplifier sensitivity.
The following screen is visible:
Use key 1 change the sensitivity. Confirm by pressing key 3 for >2 sec.
Up
Down Confirm
The following screen is visible: