ES5392.1 High Current Switch Board (6-CH) - User’s Guide
11
ETAS
Introduction
Approved Cables
The signal lines must not exceed a maximum length of 3 m.
For connecting lab power supplies and connecting an ES4408.1, use only cables
that are approved by ETAS. Appropriate cables are specified in this User’s Guide
in the chapter "Accessories" on page 29.
When ordering a power supply from ETAS (e.g. SM52-AR-60 from Delta Elek-
tronika), the matching cable is included in the scope of supply.
A list of the power supplies supported by the LCO software is located in the RTC
User's Guide in the section of the ES5392 description.
Requirements for the Installation Location
Requirements for Ventilation
Transport and Installation
To avoid damages to the hardware from electrostatic discharge, please observe
the following precautionary measures:
WARNING!
Fire hazard!
Use only approved cables for creating cable assemblies (e.g. for con-
necting the ECU and external loads). The cables used must be suitable
particularly for occurring currents, voltages and temperatures and
flame-retardant in accordance with one of the following standards
IEC 60332-1-2, IEC 60332-2-2, UL 2556/UL1581VW-1!
WARNING!
This is class A equipment. This equipment can cause radio interfer-
ence in residential areas. Should that be the case, the operator may
be requested to institute reasonable measures.
CAUTION!
The air circulation inside the ES5300.1-A Housing and the
ES5300.1-B Housing can be ensured only if all free slots are covered
with front plates. Otherwise, it may lead to overtemperatures and trip
the overtemperature protection of the ES5300.1-A or the ES5300.1-
B. For this reason, install front plates in all free slots!
CAUTION!
Some components of the ES5392.1 can be damaged or destroyed by
electrostatic discharges. Leave the plug-in card in its transport pack-
aging until its installation.
The ES5392.1 may be removed from the transport packaging, config-
ured and installed only at a workplace that is secured against electro-
static discharges.