73
9
Rewriting procedures of the Software of APH 550
The rewriting procedures of the software of APH550 are described below.
1. Preparation
・
PC
・
RS-232C straight cable
・
Software rewriting application (ComUpdate.exe)
2. Procedures
(1) Start up the application of software rewriting (ComUpdate.exe).
(2) Communication setting of the application
Set the communication port and communication speed of the application of software
rewriting.
・
Set the communication port depending on the PC environment.
・
Set the “Product” mode “APH550”.
・
Set the communication speed same with the one of APH550.
(Refer to the procedure (6) about the setting of the communication speed of APH550).
・
Set the “Hyper” mode “On”.
Setting of the communication port setting
Setting of the model
Setting of the communication speed
Setting of the Hyper mode, On and Off