![Essilor Instruments APH 550 Maintenance Manual Download Page 54](http://html1.mh-extra.com/html/essilor-instruments/aph-550/aph-550_maintenance-manual_2429049054.webp)
52
8-1 Head
8-1-1 Removal of LD case assy, Face shield, head rest, cover for hiding
screw, Head rest adjustment knob assy, near point chart fixing
knob assy
1) Remove 2 covers for hiding screw by using a thin pin etc.
2) Remove the PD case F after unscrewing 2 screws for fixing the case in the covers for
hiding screw.
3) In case of removing the right LD case F, unscrew 2 screws for fixing the case inside of
the Face shields after removing the head rest and the Face shields.
4) The right LD case F is fixed with one internal tab. Remove the right LD case F unit
by opening the case from the diagonal side of the tab and unfixing the tab.
Face shields
Covers for hiding screw/ screws for fixing case
PD case F
Right LD case F
Head rest
Left LD case F
Screws for fixing cover
Tab
Open the case from this side.
Tab
Right LD case B
Right LD case F
Summary of Contents for APH 550