USER SECTION
CHAPTER 4 – OPERATING YOUR CABINET
4.2 MENU OPTIONS
Note: This is only applicable for SPD
The diagram below illustrates the complete heirarchy of menu options available in this cabinet. Press UP
or DOWN button to move through the various menu options. Users are strongly advised to exit the menu
options after the requisite adjustment(s) have been made so that any unathorized access to the menu
options thereafter can be prevented.
SET TIME
POST PURGE
WARM UP
VELOCITY UNIT
EXP TIMER
TEMP UNIT
SETTINGS
SET CONSTANT
ZERO SENSOR
CALIB SENSOR
CALIBRATION
NEW ADMIN PIN
NEW FAN PIN
A/F MONITOR
FILTER TYPE
RESET B/H/M
RESET DEFAULT
ADMIN
SET MODE
NORMAL
MAINTENANCE
MENU
Hierarchy of menu-options
When you are entering menu options, the alarm will sound to indicate that the
microprocessor is not monitoring the operation of the cabinet and as such will not give
airflow alarms. No further warnings will be given. Therefore, it is highly recommended that
the user set the ADMIN PIN, which will restrict unauthorized access to the menu. Please
refer to section 4.2.3.1
for further details. The default ADMIN PIN from the factory is 0000.
37