21 Changi South Street 1 Singapore 486777 Phone: +65 65420833 Fax: +65 65426920
E-mail: [email protected] Website:
www.escoglobal.com/ductless
USER SECTION
CHAPTER 1 – BASIC PRODUCT INFORMATION
•
2 electrical outlets at the back wall, providing a convenient power source for equipment to be
used in the cabinet
1.7 EASY SERVICING
We design the cabinets in such a way that the number of people required for servicing them is minimal,
besides ensuring that the downtime is as short as possible.
•
Easy filter access through the front part of the cabinet for minimized downtime
•
Comfortable grip position for filter
•
Blower maintenance on top cover access
•
Each cabinet is supplied with a convenient hand tool kit and extra fasteners
•
Blower Hour Meter to help the user monitor total cabinet usage. Every 60 hours, the control
system reminds the user to test the exhaust concentration with gas detection tube to see
whether the filter is saturated
(SPD units only).
1.8 COST SAVING
Esco uses the permanently lubricated direct drive centrifugal blowers which employ an energy-efficient
external rotor type design. This helps Esco cabinets deliver excellent performance while keeping the
operating cost low at the same time.
1.9 TESTING
Before being shipped, each individual unit is extensively tested for performance and safety and delivered
with a detailed test report and certificate of performance. Testing performed at our factory laboratory
includes:
•
Operator comfort tests: noise, light, vibration
•
Filter efficiency
•
Electrical safety to IEC61010-1
Additional units are also randomly selected on a statistical sampling basis and re-tested using research-
grade instrumentation and additional test protocols at our dedicated Product Development Laboratory.
Esco performs testing in accordance with more than 20 of the world’s most recognized standards of
local, regional and international scopes. In particular, testing in our laboratory is most frequently
conducted based on:
•
ANSI/AIHA Z9.7: 1998. Recirculation of Air from Industrial Process Exhaust System
•
AFNOR NF X 15-211. Ductless fume enclosures
•
AS/NZ 2243.9: 2003. Safety in laboratories - recirculating fume cabinets
•
BS7989: 2001. Recirculating fume cupboards
•
EN 14175. European fume cupboard standards.