21 Changi South Street 1 Singapore 486777 Phone: +65 65420833 Fax: +65 65426920
E-mail: [email protected] Website:
www.escoglobal.com/ductless
USER SECTION
CHAPTER 4 – OPERATING YOUR CABINET
Note:
When the alarm is disabled, the warm up time will be automatically set to zero and cannot
be changed. When alarm is enabled, the warm up time is changeable between 3 – 15 minutes.
4.2.3.4 FILTER TYPE
This option is to select the different filter type (Standard, Acid, Mercury, Sulphur, Halogen, Aldehyde, and
Ammonia).
Press SET button to enter the FILTER TYPE menu.
Press UP or DOWN arrow button to choose the filter type value.
4.2.3.5 RESET BLOWER HOUR METER
This option is used to reset the blower hour meter. The blower hour meter keeps a count of how long the
blower has been in operation. It can count up to a maximum of 9999 hours. The counter value can be
checked while in maintenance mode. You can even draw up the maintenance schedule – including when
to change filter - based on the reading of blower hour meter. Please refer to the SERVICE SECTION for
more details.
Press SET button to enter the RESET B/H/M menu
Press SET button to confirm the reset
4.2.3.6 RESET DEFAULT
You can reset the default settings/values for few of the parameters discussed above by choosing this
option. The features that can be reset are WARM-UP period (factory default setting: 3 minutes), POST-
PURGE period (factory default setting: 0 minute), UV TIMER (factory default setting: 0 min), EXP TIMER
(factory default setting: 00:00:00), VELOCITY UNIT (factory default setting: m/s), TEMP UNIT (factory
default setting: Celsius) ADMIN PIN (factory default setting: 0000), and FAN PIN (factory default setting:
0000). Note that the calibration settings cannot be reset as it may cause the cabinet to operate in an
unsafe manner. The hour meters cannot be reset either.
Press SET button to enter the RESET DEFAULT menu
Press SET button to confirm the reset
Display sequence to set alarm setting feature
FILTER TYPE
Standard
FILTER TYPE SET
Display sequence to set FILTER TYPE
42
Summary of Contents for ASCENT OPTI BASIC
Page 2: ......