31
Ductless
Fume
Hoods
Fuse
F1
is
blown
•
Switch
off
the
cabinet.
•
See
Component
Layout
to
locate
the
relay
board.
•
Check
Fuse
F1
on
relay
board.
See
Figure
2
‐
1
below.
•
If
fuse
F1
is
blown,
as
temporary
solution,
use
F5
(spare)
to
replace
F1.
•
If
fuse
F1
is
okay,
proceed
to
next
step.
Faulty
motor
speed
controller
•
Switch
on
the
blower
on
the
cabinet.
•
See
Component
Layout
to
locate
the
motor
speed
controller
•
Adjust
the
knob
of
the
speed
controller
clockwise
to
speed
up
the
blower,
counterclockwise
to
lower
down
the
blower
speed.
•
If
still
the
blower
doesn't
work
when
fully
clockwise
already,
switch
off
the
cabinet
and
proceed
to
next
steps
below.
•
Add
jumper
cables
between
the
2
cables
going
to
the
speed
control.
This
is
bypassing
the
speed
controller.
See
Figure
2
‐
2/2
‐
3
below.
•
Switch
on
the
cabinet.
•
If
the
fan
operates
properly,
the
speed
controller
is
defective
and
needs
to
be
replaced.
•
If
the
fan
does
not
operate,
the
problem
is
not
with
the
speed
controller.
Please
proceed
to
next
step.
Faulty
capacitor
•
Switch
the
unit
off.
•
See
Component
Layout
to
locate
the
capacitor.
•
See
Figure
2
‐
4
below,
disconnect
two
cables
of
capacitor.
•
Using
a
capacitance
meter
or
a
DVM
set
to
capacitance
measurement.
Record
the
value
between
two
capacitor
cables.
•
If
the
value
is
out
of
given
ranges
on
table
1,
replace
capacitor.
•
If
the
capacitance
is
within
the
given
ranges
proceed
to
next
step.
Warning:
The
capacitor
may
still
have
some
electrical
charges
if
the
cabinet
was
on
prior
doing
this
checking.
Wait
for
1
‐
2
minutes
and
check
the
capacitor
voltage
using
a
voltmeter
to
ensure
the
capacitor
was
discharged
(voltage
less
than
5VAC)
prior
to
disconnecting
the
capacitor.
Figure
2
‐
1
Figure
2
‐
2
Figure
2
‐
3