11
Ductless
Fume
Hoods
5.
Press
UP
and
DOWN
buttons
until
display
show
“RESET
B/H/M”.
Press
SET
button
to
confirm
6.
The
display
shows
“”READ
MANUAL”
then
“PRESS
SET”.
Press
SET
button
to
confirm
7.
Press
MENU
button
to
exit
the
menu
options
2.2.3.6
Reset
Default
User
can
reset
the
default
setting
by
choosing
this
option.
The
features
being
reset
are
warm
‐
up
period
(3
minutes),
post
‐
purge
period
(0
minute),
velocity
unit
(Metric),
temperature
unit
(Celsius),
Admin.
PIN
(0009),
and
Fan
PIN
(0001).
Note
that
the
calibration
settings
cannot
be
reset
as
it
may
cause
the
BSC
to
operate
in
an
unsafe
manner.
The
hour
meters
cannot
be
reset
either.
1.
Press
MENU
button
to
enter
the
menu
options
2.
If
requested,
enter
ADMIN
PIN
digit
by
digit
by
using
the
UP,
DOWN
and
SET
buttons
3.
The
alarm
will
sound
for
a
few
second
to
indicate
that
the
hood
is
unsafe
for
work
4.
Press
UP
and
DOWN
buttons
until
display
show
“ADMIN”.
Press
SET
button
to
confirm
5.
Press
UP
and
DOWN
buttons
until
display
show
“RESET
DEFAULT”.
Press
SET
button
to
confirm
6.
The
display
shows
“”READ
MANUAL”
then
“PRESS
SET”.
Press
SET
button
to
confirm
7.
Press
MENU
button
to
exit
the
menu
options
2.2.4
Setting
Mode
Ductless
fume
hood
has
two
working
mode,
the
default
normal
mode
which
is
used
in
a
day
to
day
activity,
and
maintenance
mode.
SET
MODE
NORMAL
MAINTENANCE
MENU
2.2.4.1
Normal
Mode
Every
time
the
hood
is
restarted,
this
mode
will
be
activated
by
default.
In
this
mode,
all
alarms
and
interlocks
are
enabled.
2.2.4.2
Maintenance
Mode
Maintenance
mode
should
only
be
accessed
by
qualified
personnel
during
maintenance.
In
this
mode,
all
alarms
are
disabled
and
all
interlocks
are
defeated.
2.3
Alarm
and
Warnings
The
warning
AIR
FAIL!
indicates
that
there
is
airflow
failure.
The
operator
should
check
if
there
is
any
obstruction
to
the
airflow,
and
correct
it
if
possible.
However,
if
the
problem
continues,
the
operator
should
stop
working
as
the
hood’s
protection
may
have
been
compromised.
Call
service
or
Esco’s
local
distributor.
Other
alarms
that
indicate
a
failure
or
an
error
in
the
BSC
system:
‐
ERR.AIRFAIL
will
be
displayed
if
the
blower
is
turned
off
while
there
is
an
airflow
failure.
‐
ERR.CALIB
will
be
displayed
if
the
airflow
velocity
sensor
is
not
yet
calibrated.