49
Ductless
Fume
Hoods
8.4.2
General
Specification
ADC
‐
3E_
ADC
‐
4E_
ADC
‐
5E_
External
Dimension
(W
x
D
x
H)
1035
x
736
x
1485
mm
40.7”
x
29.0”
x
58.5”
1340
x
736
x
1485
mm
52.8”
x
29.0”
x
58.5”
1645
x
736
x
1485
mm
64.8”
x
29.0”
x
58.5”
Internal
Work
Area
(W
x
D
x
H)
965
x
596
x
765
mm
38.0”
x
23.5”
x
30.1”
1268
x
596
x
765
mm
50.0”
x
23.5”
x
30.1”
1575
x
596
x
765
mm
62.0”
x
23.5”
x
30.1”
Standard
Filter
Pre
‐
Filter
Disposable,
non
‐
washable
polyester
fiber,
85%
arrestance,
EU3
rated
Main
Filter
Activated
carbon
with
granular
media
bed
(7
different
filter
types
available,
codes
A
to
G)
Backup
Filter
HEPA
filter,
typical
efficiency
of
>99.99%
at
0.3
microns,
removes
particulates
and
aerosols
Total
Weight
of
Carbon
in
the
Main
Filter
2
x
(10.3
kg)
each
2
x
(22.7
lbs.)
each
2
x
(13.9
kg)
each
2
x
(30.6
lbs.)
each
2
x
(17.4
kg)
each
2
x
(38.3
lbs.)
each
Inflow
Air
Velocity
Initial
set
point:
0.40
m/s
or
80
fpm
Air
Volume
347
m
3
/h
(205
cfm)
457
m
3
/h
(269
cfm)
567
m
3
/h
(334
cfm)
Sound
Emission
61
dB
A
62
dB
A
62
dB
A
Light
Intensity
>1115
Lux
(>104
foot
‐
candles)
>1107
Lux
(>103
foot
‐
candles)
>1082
Lux
(>99
foot
‐
candles)
Construction
Material
Main
Body
1.2
mm
(0.05")
18
gauge
electro
‐
galvanized
steel
with
white
oven
‐
baked
epoxy
‐
polyester
Isocide
antimicrobial
powder
coated
finish
Side
Wall
Tempered
glass
Work
Top
1.5
mm
(0.06")
16
gauge
stainless
steel,
type
304,
with
4B
finish
Electrical
(220
‐
240
V
AC,
50Hz,
1
Φ
)
ADC
‐
3E1
ADC
‐
4E1
ADC
‐
5D1
Max
Power
350
W
350
W
550
W
Outlet
Amp
5
A
5
A
5
A
FLA
7
A
7
A
8
A
BTU/Hr
714
714
1122
Electrical
(110
‐
120
V
AC,
50Hz,
1
Φ
)
ADC
‐
3E2
ADC
‐
4E2
ADC
‐
5D2
Max
Power
350
W
385
W
450
W
Outlet
Amp
5
A
5
A
5
A
FLA
8.5
A
8.5
A
9
A
BTU/Hr
714
785
918
Electrical
(220
‐
240
V
AC,
60Hz,
1
Φ
)
ADC
‐
3E3
ADC
‐
4E3
ADC
‐
5D3
Max
Power
350
W
350
W
550
W
Outlet
Amp
5
A
5
A
5
A
FLA
7
A
7
A
8
A
BTU/Hr
714
714
1122
Net
Weight
175
kg
(386
lbs.)
225
kg
(496
lbs.)
245
kg
(540
lbs.)
Shipping
Weight
205
kg
(452
lbs.)
261
kg
(575
lbs.)
300
kg
(661
lbs.)
Shipping
Dimension
1130
x
840
x
1750
mm
44.5"
x
33.0"
x
68.9"
1410
x
840
x
1750
mm
55.5"
x
33.0"
x
68.9"
1730
x
840
x
1750
mm
68.1"
x
33.0"
x
68.9"
Shipping
Volume
1.66
m
3
(58.6
cu.ft)
2.07
m
3
(73.1
cu.ft)
2.51
m
3
(88.6
cu.ft)
Code
Name
Application
A
Standard
All
common
laboratory
chemicals,
especially
with
organics.
When
no
specific
requirements
are
present,
or
when
more
than
one
type
of
chemical
is
used.
B
Acid
Applications
involving
sulphur
dioxide,
hydrofluoric
acid
fumes.
Removes
inorganic/organic
acid
vapors
and
fumes.
C
Mercury
Highly
effective
for
removal
of
mercury
vapor
and
compounds.
(Stable,
non
‐
volatile
mercuric
sulphide
filter
media).
D
Sulphur
Removal
of
sulphur
compounds.
E
Halogen
Removal
of
halogen
compounds
like
Chlorine,
Fluorine,
Iodine,
Bromine,
Astatine
etc.
F
Aldehyde
Formaldehyde
applications
or
when
aldehydes
are
present.
Hospital
pathology
and
endoscopy
applications.
G
Ammonia/Amines
High
performance
removal
of
ammonia/amines
by
chemisorption.
Summary of Contents for Ascent max
Page 6: ...iv...
Page 10: ...viii...
Page 22: ...12...
Page 26: ...16...
Page 60: ...50...
Page 61: ...51 Ductless Fume Hoods APPENDIX...
Page 70: ......