25
Ductless
Fume
Hoods
Chapter 6–Recertification & Calibration
Ductless
fume
hood
generally
require
re
‐
certification,
when:
•
The
hood
is
re
‐
located
•
Hood’s
performance
is
suspected
•
After
filter
or
blower
replacement
•
At
least
once
a
year
During
recertification:
•
Cabinet
airflow
velocities
and
flow
patterns
are
verified
against
the
manufacturer’s
specifications
and
relevant
international
standards
•
Filters
are
scan
‐
tested
to
ensure
they
do
not
leak
•
Operator
comfort
tests
may
be
performed
•
If
airflow
velocities
are
found
to
be
off
set
point,
adjustments
are
made
as
part
of
the
certification
process
before
final
values
are
recorded.
Airflow
alarm
calibration,
when:
•
The
cause
of
the
airflow
alarm
cannot
be
determined
•
Re
‐
certification
indicates
the
displayed
airflow
deviates
by
>0.02m/s
(4fpm)
from
actual
measured
velocities.
6.1
Certification
Flowchart