18
The
distance
from
the
aperture
to
the
aperture
of
an
opposing
fume
hood,
fume
cupboard,
etc.
should
be
in
excess
of
3
m
(10’)
You
should
not
position
the
fume
hood
in
a
location
where
there
is
an
opposing
wall
(or
other
obstruction
likely
to
affect
airflow)
within
2
m
(7’)
of
the
front
aperture.
As
with
walls,
any
large
obstruction
such
as
a
pillar
or
column
projecting
beyond
the
plane
of
the
front
aperture
should
not
be
within
30
cm
(1’)
of
the
sides
of
the
fume
hood.
4.1.1.3
Relocating
the
Hood
Normally
ductless
fume
hood
are
rarely
moved
once
they
are
in
their
ideal
positions,
but
if
the
need
arise
to
relocate
or
even
repackage
the
unit,
then
here
is
some
considerations:
•
Before
moving
the
fume
hood,
remember
to
secure
all
moving
parts,
ex:
sash.
•
Fume
hood
is
heavy,
be
careful
with
the
weight
distribution.
Usually
it
will
take
6
or
more
people
to
move
it.
For
repackaging:
•
Bolt
the
fume
hood
to
the
pallet.
•
Strap
the
fume
hood
body
down
to
the
pallet.
•
Repackage
as
necessary.
If
possible,
use
original
packaging.
•
When
moving
the
fume
hood,
use
material
handling
equipment
and
lift
the
pallet.
4.1.2
Environmental
Requirements
•
Indoor
use
only
•
Altitude
of
up
to
2000
meter
(6600’)
•
Relative
humidity
between
20%
–
90%
•
Temperature
between
18
o
C
–
30
o
C
(65
o
F
–
86
o
F)
4.1.3
Support
Requirements
•
The
support/cabinetry
should
be
leveled
•
The
support/cabinetry
must
be
able
to
withstand
the
weight
of
the
cabinet
and
any
apparatus
within
•
The
support/cabinetry
must
be
ergonomics
•
Esco
recommends
Esco
support
stand
with
leveling
feet
for
BSC
•
WithoutEsco’s
support
stand,
Esco
cannot
guarantee
the
cabinet’s
resistance
against
tipping
and
hence
the
user
would
be
solely
responsible
for
ensuring
that
the
cabinet
is
securely
fastened
to
third
party
support/cabinetry.
•
The
use
of
non
‐
leveling
feet
Esco
support
stand
will
nullify
the
third
party
certification
(NSF
or
TÜV)
that
the
cabinet
may
have,
because
only
Esco
leveling
feet
support
stand
was
used
during
certification.
The
maximum
NSF
approved
leg
levelers
adjustment
is
50
mm
(2”).