unit is used. Use sample port-6 for validation.
Remember not to open any lid at least 15 min before starting the test nor during
the testing itself.
Hook up the gas analyzer inlet tube to the sample port. Make sure that the fit is perfect
and that no air can enter or exit the system.
The gas analyzer must have a gas return port connected to the multiroom IVF incubator
(i.e., another compartment). Only measure while the value on the gas analyzer stabilizes.
Please refer to the "13.5.1.2 CO
2
/O
2
calibration" section for more information on how to
perform the CO
2
gas calibration.
PASS: CO
2
concentration measured must not deviate more than ± 0.2% from the
setpoint.
35.7 O
2
gas concentration check
The O
2
gas concentration is checked for deviation. The gas sample port on the side of the
unit is used. Use sample port-6 for validation.
Remember not to open any lid at least 10 min before starting the test nor during
the testing itself.
Hook up the gas analyzer inlet tube to the sample port. Make sure that the fit is perfect
and that no air can enter or exit the system.
The gas analyzer must have a gas return port connected to the multiroom IVF incubator
(i.e., another compartment). Only measure while the value on the gas analyzer stabilizes.
Please refer to the "13.5.1.2 CO
2
/O
2
calibration" section for more information on how to
perform the CO
2
gas calibration.
PASS: O
2
concentration measured must not deviate more than ± 0.2% from the
setpoint.
35.8 Temperature check: Compartment bottoms
The first part of the temperature check is performed using a thermometer with a sensor
suitable for measuring temperature in a droplet of media covered with Paraffin oil, with
a resolution of 0.1 °C as a minimum.