MIRI
®
TL family`s multiroom IVF incubators User Manual Rev. 1.0
30
2.
The new setpoint page will appear. Press the "Gas Setpoint" button:
Figure 13.12
Setpoint page view in the MIRI
®
TL6 multiroom IVF incubator
3.
Use arrow up and down buttons to set the value: the arrow "DOWN" decreases the
value, the arrow "UP" increases the value. One-click changes the value by 0.1 °C.
Figure 13.13
Gas setpoint view in the MIRI
®
TL6 multiroom IVF incubator
When the desired value is set, press the "EXIT" button. The value will be saved
automatically.
13.4 O
2
setpoint
The MIRI
®
TL6 multiroom IVF incubator user interface images will be used as
an example for all the O
2
setpoints.
The O
2
setpoint can be adjusted from 2.0% to 20.0% in the MIRI
®
TL6 multiroom IVF
incubator and in the range from 5.0% to 20.0% in the MIRI
®
TL12 multiroom IVF
incubator.
The default CO
2
setpoint is 5.0%.
To change the setpoint for O
2
concentration, follow these instructions: