TL family`s multiroom IVF incubators User Manual Rev. 1.0
110
•
Always use the original filter (contact Esco Medical or your local distributor for
more details or ordering).
•
Change the filter once every year.
•
Failure to change the filter on time will result in low/no cleaning of incoming N
2
gas.
•
Warranty void if wrong/ non-original filter is used.
Please refer to the service manual for replacement instructions.
37.15 Firmware update
If Esco Medical has released a newer version of the firmware, this should be installed on
the MIRI
®
TL6 or MIRI
®
TL12 multiroom IVF incubator during the yearly scheduled
service.
Please refer to the service manual for instructions on how to update the firmware.
37.16 Software update
If Esco Medical has released a newer version of the software, this should be installed on
the MIRI
®
TL6 or MIRI
®
TL12 multiroom IVF incubator during the yearly scheduled
service.
Please refer to the service manual for instructions on how to update the software.
38 The Installation guide
This section describes when and how to install the MIRI
®
TL6 or MIRI TL12 multiroom
IVF incubator in the IVF clinic.
38.1 Responsibilities
All technicians or embryologists installing the MIRI
®
TL6 or MIRI
®
TL12 multiroom IVF
incubator must identify problems and perform any necessary calibrations, adjustments
and maintenance.
Installation personnel performing MEA (Mouse Embryo Assay) must be thoroughly
familiar with the MEA and all functions of the instrument, calibration and testing
procedures, and instruments used in the instrument's testing. MEA test is a supplemental
installation test and is not mandatory.
All individuals who will perform installation, repair and/or maintenance of the
instrument must be trained by Esco Medical or at a qualified training center. Experienced
service technicians or embryologists conduct training to ensure that the installation