8 Important safety instructions and warnings
8.1 Before installation
1.
Do not use the product if the package is damaged. Contact Esco Medical or the local
representative.
2.
Read the user manual thoroughly before use.
3.
Always keep these instructions easily accessible near the device.
8.2 During installation
1.
Never place this unit on top of other equipment that gives off heat.
2.
Place this unit on a flat, hard and stable surface.
3.
Do not place the unit on a carpet or similar surfaces.
4.
Do not defeat the safety purpose of the grounding-type (earthing) plug.
5.
A grounding-type (earthing) plug with two blades and a third prong are provided
for your safety. If the provided plug does not fit into your outlet, consult an electri-
cian to replace the outlet.
6.
Always connect the power cord to a properly grounded outlet and only use the
cord that came with the device.
7.
Do not install near any heat sources such as radiators, heat registers, stoves or
other apparatus that produce heat.
8.
Do not use this device near water sources.
9.
Use only 100% concentration CO
2
and 100% concentration N
2
gases.
10.
Always use an external HEPA filter for input CO
2
and N
2
gases.
11.
Do not use this product if the room temperature exceeds 30 °C.
12.
Place this unit in a location with adequate ventilation to prevent internal heat
build-up. Leave at least 10 cm clearance from the rear, 30 cm from the top and 20
cm from left and right to prevent overheating and allow access to the ON/OFF
switch in the back.
13.
This unit is intended for indoor purposes only.
14.
The unit must be connected to a suitable uninterrupted power supply (UPS)
source.
8.3 Post installation
1.
Refer all servicing procedures to qualified service personnel.
2.
Servicing is required according to the service manual as well as cases when the
device has been damaged in any way, e. g. suppose the apparatus has been
dropped, exposed to rain or moisture or does not operate normally. The MIRI
®
TL6
and MIRI
®
TL12 multiroom IVF incubators contain high voltage components that
may be hazardous.
3.
Unplug this device during lightning storms or when unused for an extended period
of time.