P E R F O R M I N G A N E X A M
MyLab
- G E T T I N G S T A R T E D
7 - 8
Before archival, Patient Data can be made anonymous by checking the
ANONYMIZE
box.
NOTE
The native format of the exam can not be made anonymous.
The exam can be simultaneously exported to the local archive and to external
media (in native, DICOM and multimedia formats). Check all the
destinations you want, then press
O K
to confirm and close the exam archiving
it in the selected destination(s). The system automatically shows the window
allowing to start the exam.
NOTE
At power-up, the system prompts the operator to archive the last exam
performed if the system was switched off without first closing the exam in
progress.
The exams that have been performed and not archived into the local
database can be locally saved at a later time from Archive Review. Refer to
the specific section of the “Advanced Operations” manual for further
information.
If the
PAUSE EXAM
option is checked in Saving Option Menu, pressing
EN D
EXAM
also
PA USE
will be prompted.
NOTE
At switch off, the system will inform if there are any paused exams.
Monitor adjustments
Tapping the gear icon in the advanced level of the touchscreen, you can
access a menu where quickly and intuitively adjust the brightness and contrast
of the monitor depending on your work environment. Three preset levels are
available:
DARK ROOM
,
MEDIUM DARK ROOM
, and
BRIGHT ROOM
.
You can also customize the
CONTRAST
,
BRIGHTNESS
and
GAMMA
values by
tapping
USER-DEFINED
and rotating the related knobs. Tap
FACTORY
DEFAULT
to go back to the default values.
PHYSICAL CLONE
and
SMART CLONE
act on the resolution when a second
monitor is present; refer to Advanced Operations manual for further
information.
Tap the gear icon again to exit the monitor adjustments menu.
Summary of Contents for MyLab 9VET
Page 1: ...Rev 05 February 2020 MyLab 9VET GETTING STARTED 350041300 ...
Page 2: ......
Page 5: ...MyLab G E T T I N G S T A R T E D v GETTING STARTED EC Declaration of Conformity ...
Page 6: ...MyLab G E T T I N G S T A R T E D vi RED Declaration of Conformity ...
Page 7: ...MyLab G E T T I N G S T A R T E D vii GETTING STARTED ...
Page 8: ...MyLab G E T T I N G S T A R T E D viii ...
Page 14: ...I N D E X MyLab G E T T I N G S T A R T E D xiv ...
Page 46: ...S Y S T E M O V E R V I E W MyLab G E T T I N G S T A R T E D 3 12 ...
Page 116: ...T E C H N I C A L S P E C I F I C A T I O N S MyLab G E T T I N G S T A R T E D 9 12 ...