P R E P A R I N G T H E S Y S T E M
MyLab
- G E T T I N G S T A R T E D
4 - 7
GETTI
NG STARTED
8. Switch the printer on.
WARNING
Always power any USB device (such as USB printers or external USB
archiving devices) through the trolley.
Now the system can be connected to the mains and the entire configuration
can be powered through the main switch.
CAUTION
Before connecting the peripheral verify not to exceed the maximum power
consumption limits indicated for insulated sockets. There is a risk of
blowing the system fuses.
WARNING
The maximum current supplied by the
MyLab
USB ports is 500mA (for USB
2.0) and 1A (for USB 3.0). Peripherals exceeding this limit can be connected
only if powered by their external power supply through a medical insulation
transformer.
Auxiliary Monitor
Any auxiliary monitor connected to the Display Port has not to be used for
diagnostic purposes.
NOTE
The resolution of the auxiliary monitor cannot be lower than the main
display. The system automatically shuts down whenever a lower resolution
is detected.
Monitor Connection
Connect the monitor cable to the suitable connector on
MyLab
.
Gel warmer
Gel Warmers can be mounted on
MyLab
.
WARNING
To provide required separation from AC mains for both the system and the
auxiliary devices, any gel warmer must be powered through the
MyLab
insulation transformer.
Summary of Contents for MyLab 9VET
Page 1: ...Rev 05 February 2020 MyLab 9VET GETTING STARTED 350041300 ...
Page 2: ......
Page 5: ...MyLab G E T T I N G S T A R T E D v GETTING STARTED EC Declaration of Conformity ...
Page 6: ...MyLab G E T T I N G S T A R T E D vi RED Declaration of Conformity ...
Page 7: ...MyLab G E T T I N G S T A R T E D vii GETTING STARTED ...
Page 8: ...MyLab G E T T I N G S T A R T E D viii ...
Page 14: ...I N D E X MyLab G E T T I N G S T A R T E D xiv ...
Page 46: ...S Y S T E M O V E R V I E W MyLab G E T T I N G S T A R T E D 3 12 ...
Page 116: ...T E C H N I C A L S P E C I F I C A T I O N S MyLab G E T T I N G S T A R T E D 9 12 ...