A D D I T I O N A L I N F O R M A T I O N O N S A F E T Y
MyLab
- G E T T I N G S T A R T E D
2 - 2
Electromagnetic Compatibility
This system was designed for use in the electromagnetic environments
declared in the tables below, in compliance with standard IEC 60601-1-
2:2014 (4
th
edition). The operator must make sure that s/he uses it in keeping
with this standard.
NOTE
Cables and accessories other than the supplied ones could negatively affect
EMC performance of the system.
Electromagnetic Emissions
NOTE
The EMISSIONS characteristics of this equipment make it suitable for use
in industrial areas and hospitals (CISPR 11 class A). If it is used in a
residential environment (for which CISPR 11 class B is normally required)
this equipment might not offer adequate protection to radio-frequency
communication services. The user might need to take mitigation measures,
such as relocating or re-orienting the equipment.
WARNING
Use of this equipment adjacent to or stacked with other equipment should
be avoided because it could result in improper operation. If such use is
necessary, this equipment and the other equipment should be observed to
verify that they are operating normally.
Guidance and manufacturer’s declaration – Electromagnetic emissions
MyLab
is suitable for use in the specified electromagnetic environment.
The purchaser or user of
MyLab
should assure that it is used in an electromagnetic
environment as described below:
Emission Test
Compliance
Electromagnetic Environment
RF emissions
CISPR 11
Group 1
This
MyLab
uses RF energy only for its
internal function. Therefore, the RF
emission is very low and not likely to cause
any interference in nearby electronic
equipment.
RF emissions
CISPR 11
Class A
MyLab
is suitable for use in all
establishments, other than domestic and
those directly connected to the public low
voltage power supply network that supplies
buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Complies
Class A
Voltage fluctuations
and flicker emissions
IEC 61000-3-3
Complies
Summary of Contents for MyLab 9VET
Page 1: ...Rev 05 February 2020 MyLab 9VET GETTING STARTED 350041300 ...
Page 2: ......
Page 5: ...MyLab G E T T I N G S T A R T E D v GETTING STARTED EC Declaration of Conformity ...
Page 6: ...MyLab G E T T I N G S T A R T E D vi RED Declaration of Conformity ...
Page 7: ...MyLab G E T T I N G S T A R T E D vii GETTING STARTED ...
Page 8: ...MyLab G E T T I N G S T A R T E D viii ...
Page 14: ...I N D E X MyLab G E T T I N G S T A R T E D xiv ...
Page 46: ...S Y S T E M O V E R V I E W MyLab G E T T I N G S T A R T E D 3 12 ...
Page 116: ...T E C H N I C A L S P E C I F I C A T I O N S MyLab G E T T I N G S T A R T E D 9 12 ...