![Esaote MyLab 9VET Getting Started Download Page 102](http://html.mh-extra.com/html/esaote/mylab-9vet/mylab-9vet_getting-started_2425214102.webp)
S Y S T E M M A I N T E N A N C E
MyLab
- G E T T I N G S T A R T E D
8 - 4
Check the instructions supplied by the manufacturer of the cleaning agents
for possible stricter limitation. Do not use hot cleaning agents for cleaning
the equipment.
CAUTION
Apply the cleaning agents only for the time necessary to remove the dirt
without exceeding over.
A visual inspection of the parts subjected to the cleaning process in order to
evaluate possible damages or deteriorations.
Cleaning control panel and system
To clean the control panel and system, switch the
MyLab
off, unplug it and use
a soft cloth slightly dampened with water.
If necessary use the suggested wipes, or a soft cloth slightly dampened with
one of the suggested cleaning agents.
Otherwise, apply a small amount of ammonia-free and not abrasive detergent
on a clean, soft cloth and then wipe the surface.
WARNING
Make sure that the detergent has completely evaporated before turning the
equipment on.
CAUTION
Do not use any type of ammonia- or benzene-based cleaners on the case.
Cleaning the trackball
The trackball can be accessed, for cleaning purpose only, by rotating
counterclockwise the upper locking disk.
Once the disk has been removed, clean the trackball using a soft dry cloth.
Clean the trackball housing using a cotton swab.
CAUTION
When cleaning the trackball housing, make sure not to spray any liquid into
the trackball housing.
Clean the ball rotating it in its socket. Do not remove the ball from the socket.
The X-series trackballs contain two 855 nm semiconductor lasers. The device
is designed such that it fulfills laser safety class 1M regulations, according to
IEC 60825-1 2007 2nd edition.
The laser radiation emitted is invisible to the human eye.
Summary of Contents for MyLab 9VET
Page 1: ...Rev 05 February 2020 MyLab 9VET GETTING STARTED 350041300 ...
Page 2: ......
Page 5: ...MyLab G E T T I N G S T A R T E D v GETTING STARTED EC Declaration of Conformity ...
Page 6: ...MyLab G E T T I N G S T A R T E D vi RED Declaration of Conformity ...
Page 7: ...MyLab G E T T I N G S T A R T E D vii GETTING STARTED ...
Page 8: ...MyLab G E T T I N G S T A R T E D viii ...
Page 14: ...I N D E X MyLab G E T T I N G S T A R T E D xiv ...
Page 46: ...S Y S T E M O V E R V I E W MyLab G E T T I N G S T A R T E D 3 12 ...
Page 116: ...T E C H N I C A L S P E C I F I C A T I O N S MyLab G E T T I N G S T A R T E D 9 12 ...