6.2 Reprocessing
limitation
Frequent reconditioning can damage this product. Observe the
safety instructions for checking the product. With visible damage
or functional impairment, the product may no longer be used.
6.3 Recommended equipment / substances
The service life of the product, cleaning action and sterility are
validated for the following cleaning and preparation procedures
if the recommended or equivalent substances are used in accor-
dance with the manufacturer's instructions.
•
Machine cleaning and disinfection in a G 7836 CD disinfector
(Miele & Cie. KG, Gütersloh, Germany) with the cleaning
agent neodisher ® mediclean forte (Dr. Weigert GmbH & Co.
KG, Hamburg, Germany) with the recommended washing
program.
6.4 Recommended
methods
Erbe recommends the preparation procedures described below.
Equivalent different procedures are possible if not explicitly ex-
cluded. It is incumbent on the user to ensure the suitability of the
actual procedures used by means of suitable measures (e.g. val-
idation, routine monitoring, check of material compatibility).
6.5 Dismantling
1. Remove the HF electrode from the handle.
2. Detach the connecting cable from the handle.
OBJ_DOKU-188260-006.fm Seite 19 Dienstag, 22. November 2016 8:59 08
Summary of Contents for 20190-041
Page 2: ...OBJ_DOKU 188260 006 fm Seite 2 Dienstag 22 November 2016 8 59 08...
Page 22: ...22 OBJ_DOKU 188260 006 fm Seite 22 Dienstag 22 November 2016 8 59 08...
Page 63: ...EL 20190 041 OBJ_DOKU 188260 006 fm Seite 63 Dienstag 22 November 2016 8 59 08...
Page 74: ...74 OBJ_DOKU 188260 006 fm Seite 74 Dienstag 22 November 2016 8 59 08...
Page 145: ...RU 20190 041 OBJ_DOKU 188260 006 fm Seite 145 Dienstag 22 November 2016 8 59 08...
Page 174: ...174 OBJ_DOKU 188260 006 fm Seite 174 Dienstag 22 November 2016 8 59 08...
Page 175: ...OBJ_DOKU 188260 006 fm Seite 175 Dienstag 22 November 2016 8 59 08...