5.4 Connect up the product
Connect the product to a monopolar receptacle of the electrosur-
gical unit.
5.5 Using the product
Footswitches are used to activate the product.
The electrodes can be rotated by pulling on the trigger. The more
you pull on the trigger, the greater the rotary movement. The
maximum rotation is 360 °.
6
Cleaning, disinfection, sterilization
6.1 Safety
instructions
Not suitable for manual cleaning and disinfection.
Not suitable for cleaning using an ultrasonic bath.
Never use sharp objects for cleaning.
Disinfectants must be rinsed off thoroughly after use.
Maximum air pressure during drying: 2 bar.
Maximum water pressure during rinsing: 2 bar.
During machine cleaning and disinfection, do not exceed 95 °C.
It is essential to follow the cleaning agent and disinfectant man-
ufacturers’ instructions.
During sterilization, do not exceed 134 °C.
Do not sterilize in hot air.