6.9 Check
1. Check the product for visible damage and wear:
– Damage to the product, e.g. cracks, rough surface, splin-
tering, discoloration.
– Damage to the product insulation and/or cable/connec-
tor insulation, e.g. cracks and breaks.
2. Check the mechanical functionality of the product:
– Can the trigger be pulled fully?
If damaged, do not use this product!
6.10 Packaging
1. Pack the product in disposable sterilization packaging (single
or double packaging) made from paper/polyethylene and/
or in a sterilization container.
6.11 Sterilization
Only sterilize products that have been cleaned and disinfected.
Erbe Elektromedizin recommends steam sterilization with the
method described below. If other sterilization methods are used,
Erbe Elektromedizin shall bear no responsibility.
Steam sterilization
•
Fractionated vacuum method with adequate product drying
•
Holding time 3 to 18 minutes at 132 to 134 °C
•
Sterilizers in accordance with applicable national standards
and regulations (e.g. DIN EN 13060 or DIN EN 285)
•
Sterilization process validated according to DIN EN
ISO 17665
Please follow the recommendations issued by the sterilizer man-
ufacturer concerning loading, handling, and drying times.
OBJ_DOKU-188260-006.fm Seite 21 Dienstag, 22. November 2016 8:59 08
Summary of Contents for 20190-041
Page 2: ...OBJ_DOKU 188260 006 fm Seite 2 Dienstag 22 November 2016 8 59 08...
Page 22: ...22 OBJ_DOKU 188260 006 fm Seite 22 Dienstag 22 November 2016 8 59 08...
Page 63: ...EL 20190 041 OBJ_DOKU 188260 006 fm Seite 63 Dienstag 22 November 2016 8 59 08...
Page 74: ...74 OBJ_DOKU 188260 006 fm Seite 74 Dienstag 22 November 2016 8 59 08...
Page 145: ...RU 20190 041 OBJ_DOKU 188260 006 fm Seite 145 Dienstag 22 November 2016 8 59 08...
Page 174: ...174 OBJ_DOKU 188260 006 fm Seite 174 Dienstag 22 November 2016 8 59 08...
Page 175: ...OBJ_DOKU 188260 006 fm Seite 175 Dienstag 22 November 2016 8 59 08...