Never lay this product on the patient or in his/her direct vicinity.
Damp inside the connectors may cause flash-overs that could
damage the connector or its insulation. Ensure that the product
is dried sufficiently during reconditioning and prevent fluids
from penetrating the connector during use. The older the mate-
rial, the greater the risk of unwanted damp in the connectors.
Protect this product from any form of mechanical damage! Do
not throw! Do not use force! If there is a cable, and unless stated
otherwise, do not kink it or wrap it around the accompanying
product!
Erbe Elektromedizin expressly warns against modifying the
product. Any modification exempts Erbe Elektromedizin from
any and all liability.
4
How to use
4.1 Only use cleaned and disinfected products
Clean and disinfect the product before using for the first time
and before every reuse.
4.2 Checking the product
Check the product for external damage before each use.
If damaged, do not use this product!