3. Start a tested program with the following properties:
– Sufficient cleaning effect.
– Thermal disinfection: 5 to 10 minutes at 90 to 95 °C,
A
0
≥
3000.
– Final rinse with distilled or deionized water.
– Sufficient product drying (no visible fluid residues).
4. If contaminants are still visible on the product after the end
of the routine, repeat precleaning and machine cleaning and
disinfection.
5. Remove the irrigation adapter when the program has ended.
5.8 Check
1. Check the product for visible damage and wear:
– Damage to the product, e.g. cracks, rough surface, splin-
tering, discoloration.
– Damage to the product insulation and/or cable/connec-
tor insulation, e.g. cracks and breaks.
If damaged, do not use this product!
5.9 Packaging
1. Pack the product in disposable sterilization packaging (single
or double packaging) made from paper/foil and/or in a ster-
ilization container.
5.10 Sterilization
Only sterilize products that have been cleaned and disinfected.
Erbe Elektromedizin recommends steam sterilization with the
method described below. If other sterilization methods are used,
Erbe Elektromedizin shall bear no responsibility.
OBJ_DOKU-194452-003.fm Seite 27 Donnerstag, 2. April 2020 2:47 14
Summary of Contents for 20132-249
Page 2: ...OBJ_DOKU 194452 003 fm Seite 2 Donnerstag 2 April 2020 2 47 14 ...
Page 30: ...30 OBJ_DOKU 194452 003 fm Seite 30 Donnerstag 2 April 2020 2 47 14 ...
Page 102: ...102 OBJ_DOKU 194452 003 fm Seite 102 Donnerstag 2 April 2020 2 47 14 ...
Page 216: ...216 OBJ_DOKU 194452 003 fm Seite 216 Donnerstag 2 April 2020 2 47 14 ...
Page 256: ...256 OBJ_DOKU 194452 003 fm Seite 256 Donnerstag 2 April 2020 2 47 14 ...
Page 257: ...OBJ_DOKU 194452 003 fm Seite 257 Donnerstag 2 April 2020 2 47 14 ...