Steam sterilization
•
Fractionated vacuum method with adequate product drying
•
Holding time 3 to 18 minutes at 132 to 138 °C
•
Sterilizers in accordance with applicable national standards
and regulations (e.g. DIN EN 13060 or DIN EN 285)
•
Sterilization process validated according to DIN EN
ISO 17665
Please follow the recommendations issued by the sterilizer man-
ufacturer concerning loading, handling, and drying times.
5.11 Overview of validated procedures
The following procedures have been validated as suitable for
processing the product:
•
Manual cleaning/disinfection with the detergent Cidezyme
and disinfectant Cidex OPA (Johnson & Johnson Medical
Limited, Gargrave, Skipton).
– Cidezyme: Dosage 1.6%, 5 minutes exposure time at
room temperature.
– Cidex OPA: Dosage 0.3%, 12 minutes exposure time at
20 °C.
•
Machine cleaning and disinfection in a G 7836 CD disinfector
(Miele & Cie. KG, Gütersloh, Germany) with the cleaning
agent neodisher mediclean forte (Dr. Weigert GmbH & Co.
KG, Hamburg, Germany).
– neodisher mediclean forte: Dosage 0.5%, 5 minutes ex-
posure time at 55 °C.
– Disinfection: 5 minutes at 90 °C.
•
Sterilization in an autoclave Systec V-150 (Systec GmbH,
Linden/Germany).
– Steam sterilization with saturated steam, fractionated
vacuum procedure, 3 minutes, 132 °C, 3 bar.
Erbe recommends the processing procedures described in these
instructions for use. Equivalent different procedures are possible
OBJ_DOKU-194452-003.fm Seite 28 Donnerstag, 2. April 2020 2:47 14
Summary of Contents for 20132-249
Page 2: ...OBJ_DOKU 194452 003 fm Seite 2 Donnerstag 2 April 2020 2 47 14 ...
Page 30: ...30 OBJ_DOKU 194452 003 fm Seite 30 Donnerstag 2 April 2020 2 47 14 ...
Page 102: ...102 OBJ_DOKU 194452 003 fm Seite 102 Donnerstag 2 April 2020 2 47 14 ...
Page 216: ...216 OBJ_DOKU 194452 003 fm Seite 216 Donnerstag 2 April 2020 2 47 14 ...
Page 256: ...256 OBJ_DOKU 194452 003 fm Seite 256 Donnerstag 2 April 2020 2 47 14 ...
Page 257: ...OBJ_DOKU 194452 003 fm Seite 257 Donnerstag 2 April 2020 2 47 14 ...