P a g e
|
51
Follow the same pressure calibration procedure as detailed in the previous section then tap the Analyze
button in the lower right corner of the screen.
Warning: Never apply vacuums greater than -
200”
H
2
0 or pressure greater than +20
0” H2O
to any of the
pressure sensors during calibration or measurement activities. These extreme pressures can damage
the pressure sensors within the Envision® gas analyzer.