MIR 9000
3–15
February 2000
3.3.2.5 Print-out
It enables output of the measurements (gas + analog) with the current (rate of 1 to 1440 min) on one
of the RS232 / RS422 communications ports to send to a serial printer or to capture them in a text file.
It is IMPERATIVE for the protocol installed on the communications port to be “PRN”. This protocol is
declared in the “
Configuration / Serial link”
menu.
Remark :
When the rate is 0000 the screen is displayed when a character is received on the selected
communications port:
The value is an average taken over the selected period.
Format of the screen displayed cyclically (variable size depending on the number of measurements).
DD-MM-YY HH:MM:SS < name_1 = > <data_1><unit_1>........< name_x = ><data_x><unit_x>
<CR/LF>
Example
19-05-1999 09:35:34 SO2 = 2.151 PPM NO = 5.091 mg/Nm3 CO2 = 7.523 % MX09 = 1013 mv