7. Menu structure of M-Sens WR3
This software is available for download on our website:
https://www.envea.global/solutions/process-optimization/dahs-software/
The menu structure supports the user when adjusting the measuring range, the calibration, the
measurement values and the choice of additional functions. In this connection, the numbering
both on the display and in the PC interface is identical:
1. Measuring range
Set all relevant measuring range settings
1.1 Tag No.
Input:
Name of the measurement point or
free text (10 characters)
product.
1.2 Unit
Input:
unit text, e. g. kg
Required mass flow unit.
1.3 Decimal point
Selection:
Number representation and decimal
0000, 0.000, 00.00, 000.0 point-accuracy in the measurement menu.
1.4 Set point low
Input:
0 … 9999
Throughput rates under this value will not
not be displayed at the current output.
This does not concern the display indicator,
totaliser or pulse output (= 4mA).
1.5 Set point high
Input:
0 … 9999
Throughput rates above this value will not
will not be displayed at the current output.
This does not concern the display indicator,
totaliser or pulse output (= 20mA).
1.6 Filter
Input:
0.1 s … 99.9 s
Filtering of measurement for the indicator
and the output values.