6 Commissioning
Topclean S
56
Hauser
6.6
Description of functions
6.6.1
Set up 1 – Sensor input
In this menu, you can change the measured value acquisition settings, such asthe operating mode,
the measuring principle or the electrode type.
Apart from the signal attenuation, you have already made all menu settings at the first
commissioning in First start up (see page 51). You can change the selected values in this menu.
To access the parameter setting menu, you have to enter your specialist code (see page 45, page 58).
Proceed as follows:
⇒
⇒
PARAM
pH
7.00
Param
Operating level
Read only
Operator
Specialist
Edit
Next E
Hold
pH
7.00
Param
Set up 1
Sensor input
Display
Access codes
Current output
Relays
Edit
Next E
Hold
CHOICE
(default = bold)
INFO
pH
Redox mV
Redox %
Operating mode selection
If the operating mode changes, the user settings are automatically reset.
Glass el.
Glass el.
Antimony
ISFET
7.0
4.6
Select electrode type 1
(pH only)
!
Note!
• When changing from glass or antimony electrode to ISFET, the temperature sensor is
reset to Pt 1000 as a default. Conversely, Pt 100 is selected.
• The Topclean S version glass / ISFET (CPC30-xx2xxxxxxx) is factory set for
measurement with glass electrodes.
• For digital sensors with Memosens technology, only the electrode type Glass el. 7.0 is
available.
solution ground
no solution
ground
Select connection type
solution ground
= with potential matching (PML)
no solution ground
= without PML
!
Note!
For further information, see page 20.
pH/redox:
Temperature:
(00...30s)
00 s
00 s
Set measured value attenuation
The mean value over the set time is displayed. There is no attenuation for the 00s value.
pH
7. 00
Hold
Param
pH
Edit
[¯ ]
[E]
Sensor input
Redox/ORP mV
Redox/ORP %
Next
Summary of Contents for Topclean S CPC30
Page 154: ...10 Technical data Topclean S 154 Endress Hauser...
Page 156: ...11 Appendix Topclean S 156 Endress Hauser C07 CPC30xxx 19 06 08 en 006 eps...
Page 157: ...Topclean S 11 Appendix Endress Hauser 157 C07 CPC30xxx 19 06 08 en 007 eps...
Page 158: ...11 Appendix Topclean S 158 Endress Hauser C07 CPC30xxx 19 06 08 en 008 eps...
Page 159: ...Topclean S 11 Appendix Endress Hauser 159...
Page 160: ...11 Appendix Topclean S 160 Endress Hauser C07 CPC30xxx 19 06 08 en 002 eps...
Page 161: ...Topclean S 11 Appendix Endress Hauser 161 C07 CPC30xxx 19 06 08 en 003 eps...
Page 162: ...11 Appendix Topclean S 162 Endress Hauser C07 CPC30xxx 19 06 08 en 014 eps...
Page 163: ...Topclean S 11 Appendix Endress Hauser 163 C07 CPC30xxx 19 06 08 en 015 eps...
Page 164: ...11 Appendix Topclean S 164 Endress Hauser C07 CPC30xxx 19 06 08 en 011 eps...
Page 165: ...Topclean S 11 Appendix Endress Hauser 165 C07 CPC300xx 19 06 08 en 010 eps...
Page 166: ...11 Appendix Topclean S 166 Endress Hauser C07 CPC30xxx 19 06 08 en 012 eps...
Page 167: ...Topclean S 11 Appendix Endress Hauser 167 C07 CPC30xxx 19 06 08 en 013 eps...
Page 176: ...Index Topclean S 176 Endress Hauser...
Page 178: ...www endress com worldwide BA236C 07 en 11 04 51504339 Printed in Germany FM SGML 6 0 DT...