6 Commissioning
Topclean S
64
Hauser
Temperature compensation process (pH only):
Select temp. table
Edit table
Reference temperature
Selection
Enter / activate customer-specific temperature compensation
tables.
Select temp. table
= select for activation
Select temperature table:
Medium 1
Medium 2
Medium 3
no
Select medium
No
= no medium compensation
Edit table:
Medium 1
Medium 2
Medium 3
Select medium
Medium compensation curves can be entered
as a table for three different media.
02
(2 ... 10)
Entry of the number of support points (value pairs)
Value pair: pH/redox and temperature
°C
020.0 °C
025.0 °C
pH
02.00
04.00
Value pair entry
Enter pH/redox and temperature (number of required value pairs =
number of support points entered in previous field).
OK
Delete pair(s)
Selection:
Are the value pairs OK or do you want to delete pairs?
°C
020.0 °C
025.0 °C
pH
02.00
04.00
Delete:
Select the rows to be deleted, delete them with
T
and confirm
this with "E".
Table valid
Note in display:
The table is active after confirmation by pressing "E". Cancel by
pressing "PARAM".
Reference temperature:
For laboratory measurement:
25.0 °C
(0 ... 100 °C)
Enter reference temperature
to which the medium temperature shall be compensated. Enter the
temperature at which the pH reference value of the process is
defined (e.g. the ambient temperature in the laboratory).
CHOICE
(default = bold)
INFO
Hold
pH
7. 00
[E]
Medium comp.
Next
Edit[¯]
Param
Select temp. table
Edit table
Reference temperat
Hold
pH
7. 00
[E]
Comp. table 1
Next
Edit[ ]
Param
Medium 1
Medium 2
Medium 3
No
Hold
pH
7. 00
[E]
Comp. table
Next
Edit[¯]
Param
Medium 1
Medium 2
Medium 3
Hold
pH
7. 00
[E]
Reference temp.
Next
Edit[ ¯®]
Param
Lab measurement
025.0
°C
-20.0... 150.0 °C
Summary of Contents for Topclean S CPC30
Page 154: ...10 Technical data Topclean S 154 Endress Hauser...
Page 156: ...11 Appendix Topclean S 156 Endress Hauser C07 CPC30xxx 19 06 08 en 006 eps...
Page 157: ...Topclean S 11 Appendix Endress Hauser 157 C07 CPC30xxx 19 06 08 en 007 eps...
Page 158: ...11 Appendix Topclean S 158 Endress Hauser C07 CPC30xxx 19 06 08 en 008 eps...
Page 159: ...Topclean S 11 Appendix Endress Hauser 159...
Page 160: ...11 Appendix Topclean S 160 Endress Hauser C07 CPC30xxx 19 06 08 en 002 eps...
Page 161: ...Topclean S 11 Appendix Endress Hauser 161 C07 CPC30xxx 19 06 08 en 003 eps...
Page 162: ...11 Appendix Topclean S 162 Endress Hauser C07 CPC30xxx 19 06 08 en 014 eps...
Page 163: ...Topclean S 11 Appendix Endress Hauser 163 C07 CPC30xxx 19 06 08 en 015 eps...
Page 164: ...11 Appendix Topclean S 164 Endress Hauser C07 CPC30xxx 19 06 08 en 011 eps...
Page 165: ...Topclean S 11 Appendix Endress Hauser 165 C07 CPC300xx 19 06 08 en 010 eps...
Page 166: ...11 Appendix Topclean S 166 Endress Hauser C07 CPC30xxx 19 06 08 en 012 eps...
Page 167: ...Topclean S 11 Appendix Endress Hauser 167 C07 CPC30xxx 19 06 08 en 013 eps...
Page 176: ...Index Topclean S 176 Endress Hauser...
Page 178: ...www endress com worldwide BA236C 07 en 11 04 51504339 Printed in Germany FM SGML 6 0 DT...