6 Commissioning
Topclean S
48
Hauser
6
Commissioning
6.1
Special features for measurement with digital sensors
with Memosens technology
!
Note!
Digital sensors with Memosens technology can only be used in combination with the appropriate
Topclean S version (CPC30-xx5xxxxxxx).
You can retrofit an existing Topclean S for Memosens technology (see page 134).
Commissioning
Digital sensors with Memosens technology save calibration data. Therefore, commissioning of these
sensors differs from the commissioning of standard electrodes. Proceed as follows:
1.
Install the transmitter and the assembly.
2.
Connect the transmitter and the sensor cable.
3.
Configure the transmitter according to your specific requirements (see »Description of
functions« on page 56).
4.
Connect the factory-calibrated sensor with Memosens technology and immerse it in the
medium or buffer.
5.
The saved sensor-specific calibration data are automatically transferred to the transmitter.
6.
The measured value is displayed.
Data storage
Digital sensors are able to store the following data:
• Manufacturing data
– Serial number
– Order code
– Date of manufacture
• Calibration data
– Calibration date
– Calibrated slope at 25 °C / 77 °F
– Calibrated zero point at 25 °C / 77 °F
– Temperature offset
– Serial number of the transmitter used for the last sensor calibration
– Buffer values of the last calibration
– Change in slope compared to preceding calibration
– Change in zero point compared to preceding calibration
• Application data
– Temperature application range
– pH application range
– Operating hours at temperatures above 80 °C / 176 °F and 100 °C / 212 °F
– Operating hours at very low and very high pH values (Nernst voltage below -300 mV,
above +300 mV)
– Number of sterilisations
To display these sensor data, select
Ð
➠
Ext. sensor data.
Summary of Contents for Topclean S CPC30
Page 154: ...10 Technical data Topclean S 154 Endress Hauser...
Page 156: ...11 Appendix Topclean S 156 Endress Hauser C07 CPC30xxx 19 06 08 en 006 eps...
Page 157: ...Topclean S 11 Appendix Endress Hauser 157 C07 CPC30xxx 19 06 08 en 007 eps...
Page 158: ...11 Appendix Topclean S 158 Endress Hauser C07 CPC30xxx 19 06 08 en 008 eps...
Page 159: ...Topclean S 11 Appendix Endress Hauser 159...
Page 160: ...11 Appendix Topclean S 160 Endress Hauser C07 CPC30xxx 19 06 08 en 002 eps...
Page 161: ...Topclean S 11 Appendix Endress Hauser 161 C07 CPC30xxx 19 06 08 en 003 eps...
Page 162: ...11 Appendix Topclean S 162 Endress Hauser C07 CPC30xxx 19 06 08 en 014 eps...
Page 163: ...Topclean S 11 Appendix Endress Hauser 163 C07 CPC30xxx 19 06 08 en 015 eps...
Page 164: ...11 Appendix Topclean S 164 Endress Hauser C07 CPC30xxx 19 06 08 en 011 eps...
Page 165: ...Topclean S 11 Appendix Endress Hauser 165 C07 CPC300xx 19 06 08 en 010 eps...
Page 166: ...11 Appendix Topclean S 166 Endress Hauser C07 CPC30xxx 19 06 08 en 012 eps...
Page 167: ...Topclean S 11 Appendix Endress Hauser 167 C07 CPC30xxx 19 06 08 en 013 eps...
Page 176: ...Index Topclean S 176 Endress Hauser...
Page 178: ...www endress com worldwide BA236C 07 en 11 04 51504339 Printed in Germany FM SGML 6 0 DT...