Topclean S
4 Wiring and pneumatic connection of Topclean S
Hauser
25
4.1.4
Digital pH sensors with Memosens technology
Measuring cable
To connect digital sensors with Memosens technology to Mycom S CPM153, you require the
CYK10 Memosens data cable with 2x2 wires, twisted pair, screen and PVC sheath.
C07-CPM2x3xx-04-06-00-xx-003.eps
ƒig. 20:
Design of CYK10 Memosens data cable
1
Coupling with integrated electronics for connection to the sensor
Connecting the sensor
C07-CPM153xx-04-06-00-xx-015.eps
ƒig. 21:
CPS11D connection with CYK10
Connect the wires as follows:
Wire
Mycom terminal
Yellow (YE)
Terminal 97
Green (GN)
Terminal 96
White (WH)
Terminal 88
Brown (BN)
Terminal 87
Screen
Grounded via metal cable gland
YE
GN
BN
WH
+ 87
- 88
Com A 96
Com B 97
1
Summary of Contents for Topclean S CPC30
Page 154: ...10 Technical data Topclean S 154 Endress Hauser...
Page 156: ...11 Appendix Topclean S 156 Endress Hauser C07 CPC30xxx 19 06 08 en 006 eps...
Page 157: ...Topclean S 11 Appendix Endress Hauser 157 C07 CPC30xxx 19 06 08 en 007 eps...
Page 158: ...11 Appendix Topclean S 158 Endress Hauser C07 CPC30xxx 19 06 08 en 008 eps...
Page 159: ...Topclean S 11 Appendix Endress Hauser 159...
Page 160: ...11 Appendix Topclean S 160 Endress Hauser C07 CPC30xxx 19 06 08 en 002 eps...
Page 161: ...Topclean S 11 Appendix Endress Hauser 161 C07 CPC30xxx 19 06 08 en 003 eps...
Page 162: ...11 Appendix Topclean S 162 Endress Hauser C07 CPC30xxx 19 06 08 en 014 eps...
Page 163: ...Topclean S 11 Appendix Endress Hauser 163 C07 CPC30xxx 19 06 08 en 015 eps...
Page 164: ...11 Appendix Topclean S 164 Endress Hauser C07 CPC30xxx 19 06 08 en 011 eps...
Page 165: ...Topclean S 11 Appendix Endress Hauser 165 C07 CPC300xx 19 06 08 en 010 eps...
Page 166: ...11 Appendix Topclean S 166 Endress Hauser C07 CPC30xxx 19 06 08 en 012 eps...
Page 167: ...Topclean S 11 Appendix Endress Hauser 167 C07 CPC30xxx 19 06 08 en 013 eps...
Page 176: ...Index Topclean S 176 Endress Hauser...
Page 178: ...www endress com worldwide BA236C 07 en 11 04 51504339 Printed in Germany FM SGML 6 0 DT...