5 Operation
StamoSens CSM 750 / CSS 70
18
EHauser
5.2.3
"PARAMETER ENTRY" Menu
!
Note!
Some settings that can be made in the "CONFIGURATION" menu affect the defaults in
the "PARAMETER ENTRY" menu. In view of this, complete the "CONFIGURATION" menu
first during initial start-up (s. Chap. 6.2).
Option
Display
Range of adjustment
Factory settings (bold)
Info
PARAMETER ENTRY
Setting the default values for
– Measuring ranges
– Alarm values
– Cleaning
Measuring range-start
0.30 ... 50.0 1/m SAK
0.30 1/m SAK
Selectable concentration range within a calibration curve
assigned as a linear function to the analog output.
The displays (LED and LC) only show measured values
which are greater than half the measuring range start and
smaller than double the measuring range end. Measuring
range start and end also form an input limit for alarm
values.
Factory settings for the measuring ranges depending on
sensor type selected in the CONFIGURATION menu
Measuring range-end
0.30 ... 50.0 1/m SAK
50.0 1/m SAK
Alarm value A
0.30 ... 50.0 1/m SAK
10.0 1/m SAK
Concentration switching values for the alarm relays. The
switch hysteresis is 2% in each case.
Factory settings for the alarm values depending on sensor
type selected in the CONFIGURATION menu
Alarm value B
0.30 ... 50.0 1/m SAK
50.0 1/m SAK
Signal filter
0 ... 100
10
The signal filter indicates the number of light flashes per
single measurement for obtaining the arithmetic mean of
the sensor signal corresponding to the concentration value.
1st measurement
01.01.00 - 31.12.99
00:00 - 23:59
01.01.99 00:00h
Time of the 1st measurement (DD.MM.YY, time hh:mm).
After every change the instrument does not wait for the
measuring interval. If the measurement is to start
immediately, set the time in the past (01.01.96 at the
earliest).
Measuring interval
0 ... 120 min
0 min
Time between two measurements. If a measuring interval =
0 min is selected, the sensor performs a measurement and
starts a new measurement immediately after the current
measurement ends.
Cleansing interval
0 ... 720 min
1 min
Factory settings for the cleansing intervals depending on
sensor type selected in the CONFIGURATION menu. If a
cleansing interval = 1 min is selected, cleaning is done
after each measurement. Cleansing interval = 0 means
cleaning is switched off.
Cleansing period
1 ... 600 s
10 s
Factory settings for the cleansing period depending on
sensor type selected in the CONFIGURATION menu