StamoSens CSM 750 / CSS 70
6 Commissioning
Hauser
25
To carry out a one-point calibration of the measuring system, proceed as follows:
Key
Display
Leave the sensor to rest in the medium for approx. 1 hour.
Take a sample in the direct vicinity of the sensor and determine the organic substance in the laboratory.
Press the
H
key
During the measurement, press the
F
key 3 times
to display the sensor frequency associated with the measured value.
Make a note of the sensor frequency value.
Press the
H
key
Press the
W
key 2 times
Confirm with the
F
key
Press the
W
key to set the "99" code.
Confirm with the
F
key
Use the
W
key to set the number of measuring points at "1".
Confirm with the
F
key
Press the
H
key
Press the
W
key 3 times
Confirm with the
F
key
Press the
W
key to set the "99" code.
Confirm with the
F
key
Use the
W
or
V
keys to set the 1st measured value to the laboratory value.
Confirm with the
F
key
Press the
H
key
Press the
W
key 4 times
Confirm with the
F
key