6 Commissioning
StamoSens CSM 750 / CSS 70
26
EHauser
!
Note!
The above described one-point calibration is only reasonable when calibrating the
measuring unit SAK. In case of calibration on DOC, TOC, BOD and COD, please carry
out a multi-point calibration. For your calibration points use a high- concentrated effluent
sample and make reasonable degrees by diluting with drinking water. Use pure drinking
water for zero point calibration.
Press the
W
key to set the "99" code.
Confirm with the
F
key
Use the
W
or
V
keys to set the previously noted sensor frequency value.
Confirm with the
F
key
Press the
H
key
The one-point calibration is now completed.
Should you wish to carry out a multiple-point calibration, repeat the steps previously mentioned.