2-Pack Analyzer System
1
–
8
4900002256 rev. B 4-13-21
Note that contamination of the mirrors results solely in lower overall signal.
However, by tuning the laser off-resonance as well as on-resonance and
normalizing the data, the technique self calibrates every scan resulting in
measurements that are unaffected by mirror contamination.
Differential TDLAS
Similar to TDLAS, this SpectraSensors technology involves subtracting two
spectrums from one another. A “dry” spectrum, a response from the sample
when the analyte of interest has been completely removed, is subtracted from
the “wet” spectrum, a response from the sample when the analyte is present.
The remainder is a spectrum of the pure analyte. This technology is used for
very low or trace measurements and is also useful when the background matrix
changes over time. Refer to
http://www.spectrasensors.com/spectral-
subtraction-in-tdlas-analysis/
for more information.
Wavelength modulation spectroscopy (WMS) signal detection
SpectraSensors takes the fundamental absorption spectroscopy concept a step
further by using a sophisticated signal detection technique called wavelength
modulation spectroscopy (WMS). When employing WMS, the laser drive
current is modulated with a kHz sine wave as the laser is rapidly tuned. A lock-
in amplifier is then used to detect the harmonic component of the signal that
Normalized Absorption Signal
Wavelength [a.u.]
Signal [a.u.
]
Figure 1–3
Typical normalized absorption signal
from a laser diode absorption spectrometer
100
99
98
97
96
95
Y
Summary of Contents for SS1000
Page 2: ......
Page 4: ......
Page 8: ...2 Pack Analyzer System iv 4900002256 rev B 4 13 21 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 12: ...2 Pack Analyzer System viii 4900002256 rev B 4 13 21 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 66: ...2 Pack Analyzer System 4 12 4900002256 rev B 4 13 21 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 84: ...2 Pack Analyzer System A 18 4900002256 rev B 4 13 21 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 120: ...2 Pack Analyzer System Index 4 4900002256 rev B 4 13 21 This page intentionally left blank ...
Page 121: ...www spectrasensors com contact 4900002256 Rev B ...