2-Pack Analyzer System
3
–
14
4900002256 rev. B 4-13-21
12.
To complete the connections, connect the other end of the current
loop wires to a current loop receiver, the external serial cable to a
serial port on your computer and the alarm cables to the appropriate
alarm monitors.
To connect the Analyzer B signal and alarm cables
1.
Disconnect power from Analyzer B (right side) and open the
electronics enclosure cover. Take care not to disturb the electrical
assembly inside.
2.
Run conduit from the signal/alarm receiving station to the conduit
hub in the upper right-hand corner of the Analyzer B electronics
“Application of conduit lubricant”
3-10 and analyzer drawings in Appendix A.
Table 3–1
Analyzer A input/output signal connections
Terminal
Description
D-Conn
Color
1
Analyzer A Serial RX
Pin-3
Black
2
Analyzer A Serial TX
Pin-2
Red
3
COM Serial Ground
Pin-5
Shield
4
N/C
5
N/C
6
N/C
7
Analyzer A Current Loop +
8
Analyzer A Current Loop -
9
Analyzer B Current Loop +
10
Analyzer B Current Loop -
11
Digital Input
12
Digital Input
Hazardous voltage and risk of electric shock
. Be sure power
to the analyzer is turned off before opening the electronics
enclosure and making any connections.
Conduit seals should be used where appropriate in compliance
with local regulations.
Summary of Contents for SS1000
Page 2: ......
Page 4: ......
Page 8: ...2 Pack Analyzer System iv 4900002256 rev B 4 13 21 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 12: ...2 Pack Analyzer System viii 4900002256 rev B 4 13 21 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 66: ...2 Pack Analyzer System 4 12 4900002256 rev B 4 13 21 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 84: ...2 Pack Analyzer System A 18 4900002256 rev B 4 13 21 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 120: ...2 Pack Analyzer System Index 4 4900002256 rev B 4 13 21 This page intentionally left blank ...
Page 121: ...www spectrasensors com contact 4900002256 Rev B ...